General Purpose Stirred Water Baths (TC120 Model)

General Purpose Stirred Water Baths (TC120 Model)

Part No: TC120-ST38

Availability: In stock Shipping within 4 - 6 Weeks

Quick Overview

Grant TC120 series systems combine the versatile Grant Optima™ TC120 digital heating circulator with a choice of three economy plastic (P) tanks or five stainless steel (ST) tanks with different capacities.
SKU
TC120-ST38
Please Login to Account

Please Login / Register to view prices.

TC120 Series ST38 - 38L Stainless Steel

38L stainless steel tank and immrsion thermostat, ambient -15°C to 120°C

Min/Max Liquid Depths: 125/180 mm

TC120 Series 

Operating temperature range -15°C  to 120°C, stability ± 0.05°C

Versatile mid-range model with digital thermostatic control unit and stainless steel tank and a comprehensive specification to suit most applications for high precision temperature control.

Grant TC120 series systems combine the versatile Grant Optima™ TC120 digital heating circulator with a choice of three economy plastic (P) tanks or five stainless steel (ST) tanks with different capacities.

The Grant TC120 heating circulator incorporates Intelligent Control Optimisation™(ICO) for adaptive intelligent PID temperature control and includes a powerful integral pump, making TC120 systems suitable both for immersing samples and circulating temperature controlled fluid to external devices. It includes a convenient timer function for reaction timing and a variable high temperature alarm setting. The temperature range and heater power are automatically limited according to the liquid type selected.

- Simple-to-use rotor plus two keys provide access to the interactive interface for fast, accurate set-up
- Clear 4-digit display for instant reassurance - easy to read from a distance
- Visual and audible alarms alert you when your attention is required
- Operating setpoint plus 3 adjustable preset temperatures for convenience
- User calibration facility for optimum accuracy at the required operating temperature
- Integral pump - max. flow rate 16L/min
- Low liquid protection by float switch
- User adjustable overtemperature cut-out protects your samples and your workplace
- Maximum cooling/heating range (dependent on tank size): ambient +5 to 99ºC with plastic (P) tank, or -15* to 120ºC with stainless steel (ST) tank
- Stability ± 0.05ºC

 

Note: *Accessory cooling is required for operation at or below ambient temperature

 Note: Alternatively, the Grant Optima TC120 heating circulator can also be used to regulate the temperature in vessels other than Grant tanks. Such vessels must have vertical sides and a maximum wall thickness of 35 mm for rectangular tanks, 30 mm for circular tanks (300 mm diameter) and a capacity of up to 50 litres. The control unit is attached to other vessels by means of a clamp.



More Information
Lead Time 4 - 6 Weeks
Accreditation N/A
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account